Title Određivanje vrijednosti doznih indikatora za postupke planarne mamografije i digitalne tomosinteze dojke
Title (english) Determination of dose indicators for full-field mammography and digital breast tomosynthesis
Author Magdalena Zahtila
Mentor Doris Šegota Ritoša (mentor)
Mentor Slaven Jurković (komentor)
Committee member Maja Karić (predsjednik povjerenstva)
Committee member Petra Valković Zujić (član povjerenstva)
Committee member Doris Šegota Ritoša (član povjerenstva)
Committee member Slaven Jurković (član povjerenstva)
Granter University of Rijeka Faculty of Health Studies Rijeka
Defense date and country 2023-11-02, Croatia
Scientific / art field, discipline and subdiscipline BIOMEDICINE AND HEALTHCARE Clinical Medical Sciences
Abstract Periodično praćenje apsorbirane doze predane pacijentu tijekom dijagnostičkih postupaka te određivanje tipičnih vrijednosti doznih indikatora za najčešće pretrage jedan je od važnih koraka u procesu optimizacije. Za mamografske postupke u Republici Hrvatskoj dozni indikator za koji se uspostavlja tipična vrijednost je srednja glandularna doza. Dijagnostička referentna razina za srednju glandularnu dozu po jednoj projekciji je definirana za debljinu dojke 5,5 cm ± 0,5 cm za standardne mamografske postupke. Trenutno nije definirana tipična vrijednost srednje glandularne doze za digitalnu tomosintezu dojke. Klinički bolnički centar Rijeka prva je zdravstvena ustanova u Republici Hrvatskoj u kojoj je uspostavljen sustav osiguranja kvalitete te su fizičari sa Zavoda za medicinsku fiziku i zaštitu od zračenja čiji su poslovi i radni zadaci vezani za primjenu fizike u dijagnostičkoj i intervencijskoj radiologiji odredili tipične vrijednosti doznih indikatora za najčešće dijagnostičke postupke koji se temelje na primjeni snopova X-zraka. Budući da je te vrijednosti potrebno periodično revidirati i usporediti s nacionalnim referentnim razinama, u sklopu ovog rada provedena je revizija tipičnih vrijednosti srednje glandularne doze na uređaju za mamografiju Hologic Selenia Dimensions koji se koristi na Kliničkom zavodu za dijagnostičku i intervencijsku radiologiju, KBC Rijeka. Osim tipičnih vrijednosti za standardne mamografske postupke izračunate su i vrijednosti za digitalnu tomosintezu dojke. Dobivene vrijednosti su uspoređene s podacima iz recentne međunarodne literature.
Ciljevi ovog rada bili su odrediti tipične vrijednosti doznih indikatora za postupke planarne mamografije i digitalne tomosinteze dojke na Kliničkom zavodu za dijagnostičku i intervencijsku radiologiju, KBC Rijeka, utvrditi postoji li statistički značajna razlika u vrijednostima doznih indikatora za planarne mamografske postupke i digitalnu tomosintezu dojke u KBC Rijeka, usporediti dobivene tipične vrijednosti doznih indikatora za planarne mamografske postupke s vrijednostima nacionalnih dijagnostičkih referentnih razina u Republici Hrvatskoj te usporediti dobivene tipične vrijednosti doznih indikatora za digitalnu tomosintezu dojke s vrijednostima objavljenim u recentnim međunarodnim publikacijama.
Ukupan broj pacijenata kojima je učinjena mamografija u zadanom razdoblju bio je 1891. Nakon isključenja određenih grupa pacijenata analizirani su podaci za ukupno 135 žena podvrgnutih standardnom mamografskom postupku te 285 žena kojima je učinjena digitalna tomosinteza. Medijan vrijednosti srednje glandularne doze za standardnu mamografiju za debljinu komprimirane dojke 5,5 cm ± 0,5 cm u KBC Rijeka iznosi 1,51 mGy po jednoj projekciji, dok za digitalnu tomosintezu iznosi 1,99 mGy. Rezultat t-testa za nezavisne uzorke pokazao je da postoji statistički značajna razlika između srednje glandularne doze za standardnu mamografiju i digitalnu tomosintezu na razini značajnosti od 5%. Utvrđeno je da su dobivene tipične vrijednosti doznih indikatora u KBC Rijeka za planarne mamografske postupke manje od nacionalnih dijagnostičkih referentnih razina u Republici Hrvatskoj. Dobivene tipične vrijednosti doznih indikatora za digitalnu tomosintezu dojke usporedive su s podacima objavljenim u recentnim međunarodnim publikacijama. Rezultati ovog rada pokazali su da su vrijednosti srednje glandularne doze u KBC Rijeka u skladu s nacionalnim i međunarodnim preporukama.
Abstract (english) Regular monitoring of the absorbed dose delivered to the patient during diagnostic procedures and determination of typical values of dose indicators is an important step of the optimization process. In the Republic of Croatia, the dose indicator for mammography procedures is the average glandular dose. The national diagnostic reference value for average glandular dose per projection is defined for a breast thickness of 5,5 cm ± 0,5 cm for standard mammography procedures. The national diagnostic reference value for average glandular dose for digital breast tomosynthesis has not yet been defined. Clinical Hospital Center Rijeka was the first healthcare facility in the Republic of Croatia to implement a comprehensive quality assurance program and to establish typical values of dose indicators for the most common diagnostic procedures based on the use of X-ray beams. Since a regular review of typical values and comparison with national reference values is necessary, a review of typical values of mean glandular dose for the Hologic Selenia Dimensions mammography device used in the Department of Diagnostic and Interventional Radiology at Clinical Hospital Center Rijeka was performed as part of this work. In addition to typical values for standard mammography procedures, typical values for digital breast tomosynthesis were also determined. The determined values were compared with recently published data from.
The objectives of this work were: to determine the typical values of dose indicators for planar mammography procedures and digital breast tomosynthesis at the Department of Diagnostic and Interventional Radiology of Clinical Hospital Center Rijeka, to examine a statistically significance of the difference in the values of dose indicators for planar mammography procedures and digital breast tomosynthesis in the Clinical Hospital Center Rijeka, to determine whether the obtained typical values of dose indicators for planar mammographic procedures are in confidence with the national diagnostic reference values in the Republic of Croatia, and to determine whether the obtained typical values of dose indicators for digital breast tomosynthesis are comparable with the data published in recent publications.
The total number of patients in the indicated period was 1891, and after exclusion of certain groups of patients, data from a total of 135 women who underwent standard mammography and 285 women who underwent digital tomosynthesis were analyzed.
The median value, i.e., the typical value of the average glandular dose for standard mammography with a compressed breast thickness of 5,5 cm ± 0,5 cm at the Clinical Hospital Center Rijeka is 1,51 mGy per projection, while for digital tomosynthesis it is 1,99 mGy. The result of the independent samples t-test showed that there is a statistically significant difference between the mean glandular dose for standard mammography and digital tomosynthesis at the 5% significance level. It was found that the typical values of dose indicators established in the Clinical Hospital Center Rijeka for planar mammography procedures are lower than the national diagnostic reference values in the Republic of Croatia. The determined typical values of dose indicators for digital breast tomosynthesis are comparable with the values published in recent publications. The results of this work show that mammography practice in Clinical Hospital Center Rijeka is in accordance with national and international recommendations.
Keywords
digitalna tomosinteza dojke
dijagnostičke referentne razine
planarna mamografija
osiguranje kvalitete uporabe ionizirajućeg zračenja
tipične vrijednosti doznih indikatora
Keywords (english)
diagnostic reference levels
digital breast tomosythesis
mammography
quality assurance
typical values of dose indicators
Language croatian
URN:NBN urn:nbn:hr:184:376968
Study programme Title: Radiological Technology Study programme type: professional Study level: undergraduate Academic / professional title: stručni/a prvostupnik/prvostupnica (baccalaureus/baccalaurea) radiološke tehnologije (stručni/a prvostupnik/prvostupnica (baccalaureus/baccalaurea) radiološke tehnologije)
Type of resource Text
File origin Born digital
Access conditions Open access
Terms of use
Created on 2023-10-26 09:39:02